Are there guard columns available specifically for the Waters BEH OST (oligonucleotide separation technology) columns? - WKB25299

Article number: 25299

ENVIRONMENT

  • HPLC
  • UPLC
  • guard column
  • oligonucleotide
  • analytical
  • prep

ANSWER

No, there are no guard products available specifically for the BEH OST columns.

ADDITIONAL INFORMATION

Using a guard may affect the resolution.

BEH OST columns and proposed guards that can be used:

186003949 - ACQUITY UPLC OST C18 1.7µm 2.1x50mm column—use 186003975 - ACQUITY BEH C18 1.7µM VANGUARD Pre-Column
186003950 - ACQUITY UPLC OST C18 1.7µm 2.1x100mm column—use 186003975 - ACQUITY BEH C18 1.7µM VANGUARD Pre-Column
186005516 - ACQUITY UPLC OST C18 1.7µm 2.1x150mm column—use 186003975 - ACQUITY BEH C18 1.7µM VANGUARD Pre-Column

186003952 - XBridge OST C18 2.5µm 2.1x50mm column—use 186007949 - VanGuard Cartridge Holder and 186007772 - XBridge BEH C18 V-Gd Cart 2.5µ 2.1X5 XP
186003953 - XBridge OST C18 2.5µm 4.6x50mm column—use 186007949 - VanGuard Cartridge Holder and 186007774 - XBridge BEH C18 V-Gd Cart 2.5µ 3.9X5 XP

Note: The XP (eXtended Performance) hardware for VanGuard cartridges 186007772 and 186007774 has Parker style end nuts that require a shorter length of tubing from the ferrule to the tip of the tubing. If you have seated stainless steel tubing and compression screws and ferrules into a traditional Waters HPLC column, such as 186003952 or 186003953, the ferrule will have seated in order to make the tubing from the ferrule to the tip of tubing to a Waters hardware ferrule depth, which is longer than Parker style. Because this depth is too long when connecting to XP hardware (Parker style) and will need to be replaced if connecting stainless steel tubing into the XP hardware. If you are using PEEK finger-tight fittings or other fittings that have a ferrule that does not permanently seat, you will be OK.

186008212 - XBridge Oligo BEHC18 2.5µm 10x50 Prep—no guard offered. It is recommended that you filter or centrifuge samples prior to injection onto this column.

186008962 - XBridge Oligo BEHC18 2.5um 19x50 OBD Prep - no guard offered. It is recommended that you filter or centrifuge samples prior to injection onto this column.

VanGuard™ Column Protection Products Webpage: http://www.waters.com/waters/en_US/VanGuard-Column-Protection/nav.htm?cid=134832909

id25299, UPBINARY