What is the advantage of using the CSH C18 column chemistry instead of the BEH C18 column chemistry? - WKB250290
Article number: 250290
ENVIRONMENT
- Column characteristics
- Reverse phase
- Charged surface hybrid (CSH)
- Bridged ethylene hybrid (BEH)
- C18
ANSWER
Under low ionic strength, buffered mobile phases*, the CSH C18 provides better peak shape for basic compounds and increased loading capacity compared with the BEH C18 chemistry run in these types of mobile phases.
The CSH particle also exhibits more rapid column equilibration after changing the mobile phase pH.
* Mobile phases that contain formic acid and acetic acid
ADDITIONAL INFORMATION
For additional information, see Increasing Peak Capacity in Reversed-Phase Peptide Separation with Charged Surface Hybrid (CSH) C18 Columns
and Practical Applications of Charged Surface Hybrid (CSH) Technology.