How do I scale an analytical LC method to a preparative-scale LC method on the same chemistry? - WKB247723
Article number: 247723
OBJECTIVE or GOAL
Scale an analytical LC method to a larger, preparative-scale column.
ENVIRONMENT
- Method scaling
- Preparative column
PROCEDURE
- Go to the Waters On-line Preparative Columns Calculator.
- Choose any of the following buttons on the lower portion of this webpage:
- Analytical to Prep Gradient Calculator:
- Choose the "Gradient Calculator" tab to scale:
- Gradient
- Specify the details of the analytical method on the left-hand side.
- Specify the proposed prep column dimension on the right-hand side to scale a gradient.
- Flow Rate
- Injection Volume
- Mass Load (per injection)
- Gradient
- Choose the "Method Optimization" tab to calculate:
- Pressure values for different solvents
- Estimated column back pressure
- Various other calculations
- Choose the "Dwell Volume" tab:
- This webpage shows how to calculate system dwell volume.
- Choose the "Gradient Calculator" tab to scale:
- Tubing Back Pressure and Splitter Calculator:
- Choose the "Tubing Backpressure" tab to calculate tubing pressures.
- Choose the "Tubing Splitter Calculations" tab to calculate various splitter values.
- Focused Gradient Calculator:
- Choose the "UPLC to Prep focus gradient" tab to calculate:
- Optimized gradient focus to selectively elute target analyte with a slower gradient change in the area where the target peak needs the most resolution.
- If needed, specify a new flow rate for the focus gradient using the field in the "Override Focus Flow Rate" box.
- Select Language using the "Please Select Language" button on this webpage:
- English
- Chinese
- Japanese
- Choose the "Dwell Volume" tab:
- This webpage shows how to calculate system dwell volume.
- Choose the "UPLC to Prep focus gradient" tab to calculate:
- Column Back Pressures and Mass Loads Calculator:
- Specify the analytical method conditions under "Column A":
- Column Inner Diameter
- Column Length
- Particle Size
- Solvent Name (use drop-down list)
- Flow Rate
- Mass Load
- Injection Volume
- Specify the prep column dimensions under "Column B":
- Column Inner Diameter
- Column Length
- Particle Size
- Solvent Name (use drop-down list)
- Results will show:
- Scaled flow rate
- Scaled mass load
- Scaled injection volume
- Specify the analytical method conditions under "Column A":
- Analytical to Prep Gradient Calculator:
ADDITIONAL INFORMATION
Also see:
Preparative OBD Columns Wall Chart
Waters OBD Prep Column Technology White Paper
At-Column Dilution Application Notes
Developing Focused Gradients for Isolation and Purification
On Demand Webinar:
Steven M. Collier - Preparative Chromatography: Considerations for Column Selection