What is the maximum loading capacity of an Ultrahydrogel column? - WKB246391

Article number: 246391

ENVIRONMENT

  • Column characteristics
  • Loading capacity
  • GPC
  • Ultrahydrogel

ANSWER

Molecular weight      Sample concentration column (%w/v)     Injection volume per column
                                                                                            4.6-mm i.d.      7.8-mm i.d.
< 5,000                              less than 0.30 %                         20 - 50 µL       50 to 100 µL
5,000 to 25,000                  less than 0.20%                          20 - 50 µL       50 to 100 µL
25,000 to 200,000              less than 0.10%                          10 - 30 µL       50 to 100 µL
200,000 to 2,000,000         less than 0.05%                          10 - 30 µL       50 to 100 µL
> 2,000,000                        less than 0.025%                        10 - 30 µL       50 to 100 µL