What is the difference between the ACQUITY UPLC CSH C18 column and the XSelect CSH C18 column? - WKB243481

Article number: 243481

ENVIRONMENT

  • UltraPerformance Liquid Chromatography (UPLC)
  • Charged surface hybrid (CSH
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • ACQUITY UPLC CSH C18
  • XSelect CSH C18 XP
  • XSelect CSH C18

ANSWER

ACQUITY UPLC CSH C18 columns:

  • Packed with 1.7-µm particles
  • Stable up to 18,000 psi

XSelect CSH C18 XP columns for UHPLC:

  • Packed with 2.5-µm particles
  • Pressure stability as follows:
    • 2.1 mm and 3.0 mm I.D. XP are stable up to 18,000 psi 
    • 4.6 mm I.D. XP columns are stable up to 9,000 psi 

XSelect CSH C18 columns for HPLC:

  • Packed with 3.5-µm or 5-µm particles
  • Stable up to 6,000 psi

Not able to find a solution? Click here to request help.