How do I connect a Guard-Pak holder to my HPLC system? - WKB238052
Article number: 238052
OBJECTIVE or GOAL
Connect a Guard-Pak holder (WAT088141) to an HPLC system.
ENVIRONMENT
- Column connections
- Guard-Pak
PROCEDURE
- Connect tubing from the injector to the inlet of the Guard-Pak holder using V-Detail fittings.
- Connect tubing from the outlet of the Guard-Pak holder to the column inlet using V-Detail fittings.
ADDITIONAL INFORMATION
- The Guard-Pak holder is too large to fit inside some column heaters/ovens.
- In cases where it will not fit into a column heater, the holder must be kept outside of the heater, where tubing is long enough to connect to a column inside the heater.
- The Guard-Pak inlet and outlets need to use V-Detail fittings.
- If stainless steel fittings, the ferrules need to be set to Waters Ferrule Depth Setting (see examples below).
- If using PEEK Finger-tight Fittings, simply push tubing into the inlet and outlet of the the Guard-Pak holder as you tighten the PEEK finger-tight fittings to create a good connection.
Examples of proper ferrule depth:
Note: The Guard-Pak Holder Kit includes 4/ea Stainless Steel compression Screws and 4/ea Stainless Steel Ferrules but does not contain tubing.
Also see: Guard-Pak Holder and Inserts Care & Use Manual
id238052,