What is the part number for the column connector that is included with the ACQUITY UPLC Protein BEH SEC Guard Column, 200 Å, 1.7 µm, 4.6 mm x 30 mm (186005793)? - WKB237683
Article number: 237683
ENVIRONMENT
- Column connector
- Guard column
- ACQUITY
- Bridged ethylene hybrid (BEH)
- Size exclusion chromatography (SEC)
- BEH200
ANSWER
186006613 - 0.005" X 1.75" UPLC SEC Connection Tubing (2 per pack)
- The two-piece ferrules on this connector are only lightly staked onto the tubing upon receipt.
- The two-piece ferrules are permanently seated when the fittings are tightened into a column or a guard column.
ADDITIONAL INFORMATION
Waters also offers some other connectors that can be used.
Straight connector:
700009524 - Tubing Assembly, Inline Column Connection, .005 ID
- The two-piece ferrules on this connector are only lightly staked onto the tubing upon receipt.
- The two-piece ferrules are permanently seated when the fittings are tightened into a column or a guard column.
U-shaped connectors: (may be used when a 300 mm long column is used so guard fits into column heater more easily):
700009535 - Tubing Assembly, 'U' Column Connection, .004 ID
- The two-piece ferrules on this connector are not staked onto the tubing upon receipt.
- The two-piece ferrules are permanently seated when the fittings are tightened into a column or a guard column.
700009534 - Tubing Assembly, Offset 'U' Column Connection, .004 ID
- The two-piece ferrules on this connector are not staked onto the tubing upon receipt.
- The two-piece ferrules are permanently seated when the fittings are tightened into a column or a guard column.
id237683,