What are the common causes of longer than expected retention on a reversed-phase LC column? - WKB236458
Article number: 236458
ENVIRONMENT
- C18 columns
- Reverse phase
- Retention issues
- longer retention
ANSWER
Various factors are possible. See the "Additional Information" section.
ADDITIONAL INFORMATION
- New C18 columns have full end cap coverage, leading to longer retention in some cases.
- Column heater too cold:
- If a column heater is not working properly, it is possible that it is colder than expected.
- Lower temperatures can cause analytes to elute later than expected.
- Use an external thermometer to measure the actual temperature of the column heater.
- Flow rate too slow:
- If the system check valves or pumps are not working properly, or if there is a blockage in the flow path, the flow rate could be slower than expected.
- Slower flow rate causes analytes to elute later than expected.
- Collect waste into a graduated cylinder during a run, for a timed period. Verify that the volume collected is the expected volume.
- Gradient proportioning valve not working properly:
- The gradient proportioning valve (GPV) (low-pressure mixing) might be causing too little organic or too much aqueous to be delivered.
- Either scenario can cause increased retention.
- Gradient mixing (high-pressure mixing) is not working properly:
- The system may not be mixing the mobile phases properly, which may lead to too little organic or too much aqueous delivered.
- Either scenario can cause increased retention.
- Incorrect premixed mobile phase composition:
- Having too little organic (or too much aqueous) in a premixed mobile phase will result in more retention than expected.
- Incorrect mobile phase pH.
- Weakly basic analytes retain longer when neutralized in a pH that is 2 pH units or higher above their pKa.
- Weakly acidic analytes retain longer when neutralized in a pH that is 2 pH units or lower below their pKa.
- Inadvertently using a mobile phase with the wrong pH for the method can cause unexpected retention problems.
- Larger dwell volume (gradient delay):
- A system with a larger dwell volume (gradient delay) will show longer retention compared with a system with a lower dwell volume (gradient delay):
- Check valve problem:
- Sticky check valves can result in mobile phase being pumped back into the mobile phase bottle instead of into the mobile phase flow.
- This can result in an improper ratio of aqueous to organic in the mobile phase:
- Longer retention times are seen when less organic enters the mobile phase due to a check valve issue.
- This can result in an improper ratio of aqueous to organic in the mobile phase:
- Sticky check valves can result in mobile phase being pumped back into the mobile phase bottle instead of into the mobile phase flow.
- System leak:
- Check your system for a leak by running a dynamic leak test (or similar test) to ensure that there is no leak causing an incorrect mobile phase aqueous-to-organic ratio.
- There may be additional reasons for longer than expected retention times including mobile phase may have gotten old or a mobile phase modifier is degrading..
- Some pre-mixed mobile phases can partition or separate over time (such as H2O and Isopropanol), water settles on the bottom and Retention Times increase proportionately. (Note: Higher Concentrations of Salt increase the likelihood of Separation. ACN & H2O, can separate at Higher Concentrations of Salt.)
- The Fix: Allow system to mix, or use a Stir Plate, with a stir bar in the mobile phase.
id236458, checkvalve, DLT, eluent