Why is my 9.3K polystyrene standard tailing when I run my Styragel GPC column in DMF? - WKB236107
Article number: 236107
SYMPTOMS
- All peaks broader, but tailing with 9.3K polystyrene peak
ENVIRONMENT
- GPC
- Styragel column
- Polystyrene standard
- DMF solvent
CAUSE
No lithium bromide in the DMF mobile phase
FIX or WORKAROUND
- Add 50mM LiBr to the DMF mobile phase solvent.
ADDITIONAL INFORMATION
0.05M lithium bromide is added to polar solvents like DMF to reduce non-specific binding and interaction of polymers to the GPC column.
For additional information, see Frequently Asked GPC/SEC Questions: What additives are important and when should I use them?
id236107, eluent, EMPGPC, GPCV2K, UPASMFTN