What is the maximum temperature limit for XBridge BEH C18 and BEH C18 packings? - WKB23224

Article number: 23224

ENVIRONMENT

  • XBridge
  • BEH
  • C18
  • temperature
  • HPLC
  • ACQUITY
  • UPLC
  • UHPLC
  • care and use

ANSWER

Low-pH maximum temperature = 80 degrees Celsius.

High-pH maximum temperature  =  60 degrees Celsius.

ADDITIONAL INFORMATION