What column should I try if a lipid-conjugated peptide does not elute from a C18 or a C8 column? - WKB231427
Article number: 231427
ENVIRONMENT
- Column alternatives
- Lipid-conjugated peptide
ANSWER
Options to try might include:
ACQUITY Premier Protein BEH C4, 300Å
ACQUITY UPLC BEH C4, 300Å
XBridge Premier Protein BEH C4, 300Å
XBridge Protein BEH C4, 300Å
ACQUITY UPLC BEH Phenyl
XBridge BEH Phenyl
Delta-Pak C4, 100Å
Delta-Pak C4, 300Å
ADDITIONAL INFORMATION
Use of a shorter alkyl chain such as a C4 will be less retentive compared to C18 and C8. A wider, 300Å pore may result in some loss in resolution vs. a 130Å pore size.
Use of a phenyl bonded phase may result in retention similar to that of a C8 with alternate selectivity.
Only experimentation will confirm retentivity and selectivity of a given analyte on various stationary phases.
id231427,