Skip to main content
Waters

Does Waters have a column similar to the Phenomenex Gemini NX-C18? - WKB230328

Article number: 230328

ENVIRONMENT

  • UltraPerformance Liquid Chromatography (UPLC)
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • Column equivalency

ANSWER

Yes:

UPLC:

  • ACQUITY Premier BEH C18 (1.7µm)
  • ACQUITY UPLC BEH C18 (1.7µm)

UHPLC:

  • XBridge Premier BEH C18 (2.5µm)
  • XBridge BEH C18 (2.5um)

HPLC:

  • XBridge BEH C18 (2.5µm, 3.5µm, and 5µm)

ADDITIONAL INFORMATION

ACQUITY Premier and XBridge Premier columns are packed in MaxPeak HPS (High Performance Surfaces) hardware.

MaxPeak HPS columns offer extremely inert stainless steel surfaces inside the column. This will prevent electronegative, chelating agents and negatively charged analytes from adsorbing to the inherent positive surface charges present on nonmodified stainless steel surfaces.

Also see:

Hybrid Particle Technology: Part 2: BEH: White Paper

Low Adsorption HPLC Columns Based on MaxPeak High Performance Surfaces

ACQUITY UPLC and ACQUITY PREMIER Columns Web Page

MaxPeak Premier Columns Web Page

XBridge Premier 2.5 µm Columns

XBridge Columns Web Page

XBridge BEH XP Columns Web Page

Not able to find a solution? Click here to request help.