How much added back pressure does the ACQUITY In-Line Kit (205000343) typically add when installed in front of an ACQUITY UPLC 2.1 x 100 mm column (with 0.35 mL/min flow rate and water/ACN mobile phases)? - WKB228433
Article number: 228433
ENVIRONMENT
- ACQUITY In-Line Pre-Column Filter
ANSWER
Less than 200 psi
ADDITIONAL INFORMATION
This filter kit is expected to add no more than 200 psi when installed in front of an ACQUITY UPLC, 1.7µm, 2.1x100mm column under the following conditions:
Mobile Phase A: Buffer
Mobile Phase B: ACN
Initial conditions:
Flow rate 0.35 mL/min
Temp: 30 °C
Initial mobile phase: 99:1 A/B