Can I attach MaxPeak Premier hardware, VanGuard FIT columns to a Thermo Viper fitting? - WKB226627
Article number: 226627
ENVIRONMENT
- Hardware compatibility
- VanGuard FIT column
- MaxPeak HPS
- Premier hardware
- Viper fittings
ANSWER
Yes
ADDITIONAL INFORMATION
The VanGuard FIT hardware is similar in depth to the standard UPLC column hardware and an effective high-pressure seal is created with Viper fittings, following the maximum 90-degree turn suggestion per the Thermo fitting instructions. Avoid over-torquing the Viper fitting or it will deform the ferrule.
For additional information, see What are the part numbers for the fittings that connect an ACQUITY UPLC column or 2.5um eXtended Performance (XP) Column to a non-Waters UPLC system?