ANSWER

Yes

ADDITIONAL INFORMATION

The VanGuard FIT hardware is similar in depth to the standard UPLC column hardware and an effective high-pressure seal is created with Viper fittings, following the maximum 90-degree turn suggestion per the Thermo fitting instructions. Avoid over-torquing the Viper fitting or it will deform the ferrule.

For additional information, see What are the part numbers for the fittings that connect an ACQUITY UPLC column or 2.5um eXtended Performance (XP) Column to a non-Waters UPLC system?