Can I use the BioResolve RP mAb Polyphenyl column for peptide analysis? - WKB226371
Article number: 226371
ENVIRONMENT
- Column suitability
- Peptide analysis
- BioResolve RP mAb Polyphenyl
ANSWER
It is a reverse-phase column and could be used for peptides, but most peptides would have lower retention and less peak capacity on this column because of the very large pore size and the column chemistry.
ADDITIONAL INFORMATION
For additional information, see What reverse-phase columns does Waters offer for peptide analysis?