How do I start up a new Atlantis dC18 column from its storage solution? - WKB206657
Article number: 206657
OBJECTIVE or GOAL
Start up a new Atlantis dC18 column from its storage solution; introduce mobile phase to the column without damaging the column or precipitating any buffer in the mobile phase.
ENVIRONMENT
- Column care and use
- Atlantis dC18
PROCEDURE
- The column is shipped in 100% acetonitrile. Attach the column to the HPLC system with the outlet line flushing to a waste container.
- Flush the column for 10 column volumes with 50/50 acetonitrile or methanol/water.
- Connect the column to the detector.
- Flush the column with the mobile phase for 20 column volumes.
ADDITIONAL INFORMATION
For additional information, see the Atlantis T3, DC18 and HILIC Silica Columns Care and Use Manual.
id206657, eluent