How long can a column be if used with a column manager and a column heater/cooler? - WKB202648

Article number: 202648

ENVIRONMENT

  • ACQUITY UPLC Column Manager

ANSWER

You can use columns with an inner diameter of 1.0-2.1 mm and a length of 20-150 mm. If you attach a pre-filter or guard column to a 150-mm column, it will not fit in the column manager.

If you want to use a 150-300 mm column, use a 30-cm column heater/cooler.

ADDITIONAL INFORMATION

Part numbers

  • ACQUITY UPLC High-Temperature Column Heater: 186015010
  • ACQUITY UPLC Column Heater/Cooler: 176001287
  • ACQUITY UPLC Column Manager, Right inlet: 176001285
  • ACQUITY UPLC Column Manager, Left inlet: 176001286
  • ACQUITY UPLC 30-cm Column Heater Cooler: 186015011