How do I equilibrate the Carbamate Analysis 60Å, 4 µm, 3.9 mm x 150 mm column (WAT035577) on first use? - WKB200146
Article number: 200146
OBJECTIVE or GOAL
Equilibrate the Carbamate Analysis Column, 60Å, 4 µm, 3.9 mm x 150 mm, 1/pk (WAT035577) on first use.
ENVIRONMENT
- Carbamate Analysis column
- Equilibration
PROCEDURE
- Flush the column in the same mobile phase you will use for your analysis.
- Set the flow rate to 1.5 mL/min and flush for 30 minutes.
ADDITIONAL INFORMATION
This column ships in 50:50 methanol:acetonitrile.
Equilibrate with miscible solvents only.
See: Carbamate Analysis Column Care & Use Manual
id200146, eluent