What standards should be used to calibrate the XBridge BEH450 SEC 3.5 µm (186007643) for analysis of AAVs? - WKB200111
Article number: 200111
ENVIRONMENT
- Calibration standards
- Size exclusion chromatography (SEC)
- Adeno-associated virus (AAV)
- XBridge BEH450 SEC 3.5-µm column
ANSWER
The application notes use a standard of the AAV serotype of interest to calibrate the detector response and the Waters BEH450 SEC protein standard mix with L-tryptophan spiked into it to make sure the column is separating a protein mix of known molecular weights.
ADDITIONAL INFORMATION
The separation of the AAV from large multimer aggregates and lower molecular weight fragments is to determine the percentage of AAV monomer vs. smaller fragments.
For additional information, see Size-Exclusion Chromatography Analysis of Adeno-Associated Virus (AAV) Preparations Using a 450 Å Diol-Bonded BEH Column and Fluorescence Detection.