What is the flow rate range for the Waters nano columns? - WKB199312
Article number: 199312
ENVIRONMENT
- nano columns
- ACQUITY M-Class
- nanoEase M/Z
- nanoACQUITY
- flow rate
- range
- 100um
- 75um
- 150um
- 300um
- 1.0mm
- care and use
ANSWER
75-µm ID = 0.2–0.5 µL/min
100-µm ID = 0.35–1.0 µL/min
150-µm ID = 1.0–2.5 µL/min
300-µm ID = 4.0–10.0 µL/min
1.0-mm ID = 40–100 µL/min
ADDITIONAL INFORMATION
id199312, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR