Poor peak shape / peak shouldering - WKB196303

Article number: 196303

SYMPTOMS

  • Poor peak shape  
  • Peak shouldering on all peaks
  • Peak broadening
  • Lower sensitivity   
  • Pressure stable  

ENVIRONMENT

  • HPLC/UPLC columns
  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module
  • ACQUITY Arc
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC M-Class
  • nanoACQUITY UPLC
  • ACQUITY Premier
  • ACQUITY APC

CAUSE

Bad column

FIX or WORKAROUND

Replace the column.

ADDITIONAL INFORMATION