Problem getting mAb charged variant standard retention on BioResolve SCX mAb column - WKB195920
Article number: 195920
SYMPTOMS
- No retention of mAb charged variant standard on the BioResolve SCX mAb column
ENVIRONMENT
- BioResolve SCX mAb
- Charged variant standard
- Column care and use
CAUSE
The mAb charged variant standard is not optimized for benchmarking with the IonHance CX-MS pH concentrate and gradients. The eluent strength and conditions are not optimized for the mAb charged variant standard.
FIX or WORKAROUND
- The mAb charged variant standard should be run with the BioResolve CX pH concentrates and the benchmarking gradients used with the BioResolve CX pH concentrates.
- If using the IonHance CX-MS pH concentrates with the BioResolve SCX mAb column, the column should be benchmarked with the Humanized NIST mAb standard (186009125).
ADDITIONAL INFORMATION
For additional information, see BioResolve CX pH Concentrates A and B Care & Use Manual and[1]BioResolve SCX mAb Columns Care & Use Manual
or BioResolve SCX mAb column run with:
IonHance CX-MS pH Concentrates A and B Care & Use Manual and Optimizing LC-MS Separations of mAbs using a BioResolve SCX mAb Column and IonHance CX-MS pH Concentrates
