Is it OK to use BioResolve mAb SCX columns on an ACQUITY UPLC H-Class Bio system? - WKB193755

Article number: 193755

ENVIRONMENT

  • BioResolve
  • Monoclonal Antibaody (mAb)
  • Strong cation exchange (SCX)
  • ACQUITY
  • UltraPerformance Chromatography (UPLC)
  • H-Class Bio

ANSWER

Yes

ADDITIONAL INFORMATION

Because the pH range for an ACQUITY UPLC H-Class Bio system is 2-12, a pH gradient can be run on this system while using this column.

The pH range for a gradient that uses the BioResolve CX pH Concentrates A and B (when used as directed in the care and use manual) will be from 5.0 to 10.2. 

Also see:

ACQUITY UPLC H-Class and H-Class Bio System Specifications

BioResolve SCX mAb Columns Care & Use Manual 

mAb Charge Variant Standard Care & Use Manual

BioResolve CX pH Concentrates A and B Care & Use Manual