Can I attach and run a 30-mm ID preparative column on an Alliance system? - WKB193489

Article number: 193489

ENVIRONMENT

  • Prep columns
  • Alliance 2695

ANSWER

Yes, but this is very unusual given the size and weight of the column. It is not recommended.

ADDITIONAL INFORMATION

The inner diameter size versus the flow rate capability of the 2695 is not very conducive for optimal separation capability. The weight and ID of the column are not suited for attachment inside the column heater compartment.

For additional information, see Which HPLC column hardware IDs are most efficient when attached to an Alliance system? 