Are BEH300 C4 columns compatible with 100 percent aqueous mobile phase? - WKB193254
ENVIRONMENT
- UltraPerformance Liquid Chromatography (UPLC)
- Ultra High Performance Liquid Chromatography (UHPLC)
- High Performance Liquid Chromatography (HPLC)
- Bridged ethylene hybrid (BEH)
- BEH300 C4 columns
- ACQUITY
- XBridge
- water
ANSWER
Yes
ADDITIONAL INFORMATION
BEH300 C4 columns are compatible with 100% aqueous mobile phase, but there is a chance that dewetting will occur. It is recommended that you run in at least 5% organic solvent at all times if possible, to prevent the pores from dewetting (expelling this high aqueous solution). If dewetting occurs, you will see a decrease in retention times. To re-wet the packing, simply flush the column with organic solvent to re-wet the packed bed. If the aqueous mobile phase has a buffer, it is recommended that you first flush with 100% water with no buffer before flushing with organic, in order to avoid problems with salt precipitation.