How do I clean a µBondapak Cyano-RP column? - WKB125854

Article number: 125854

OBJECTIVE or GOAL

Clean a µBondapak CN-RP column.

ENVIRONMENT

  • Reverse-phase (RP)
  • µBondapak Cyano-RP
  • Column cleaning
  • Dimethylsulfoxide (DMSO)

PROCEDURE

  1. Remove the guard cartridge.
  2. Flush the column with 100% water for 10 column volumes.
  3. Inject 200 µL of 50:50 dimethylsulfoxide/water every 5 minutes until 10 column volumes of water has passed through the column.
  4. Stop injections.
  5. Flush column with 10 column volumes of 100% water.
  6. Flush column with 10 column volumes of 100% methanol.
  7. Flush column with 10 column volumes of 100% chloroform.
  8. Flush column with 10 column volumes of 100% methanol.
  9. Flush column with 10 column volumes of 100% water.
  10. Flush column with 10 column volumes of your initial mobile phase.