What is the storage solvent recommendation for the XTerra RP18 column? - WKB119781
Article number: 119781
ENVIRONMENT
- Column storage
- XTerra RP18
ANSWER
The column can remain in 20% organic for short-term storage (defined as four days or less). Remove any buffer salts if the column will sit idle, on the HPLC for 4 days or less.
The column must be stored in 100% acetonitrile for long-term storage. Flush the column for 10 column volumes with 20% organic/80% water (or 100% water) to remove buffers, and then flush the column for 15 column volumes with 100% acetonitrile.
ADDITIONAL INFORMATION
For additional information, see the XTerra and XTerra Prep Columns Care and Use Manual.