Can I use deionized water as the mobile phase on Ultrahydrogel columns? - WKB118128
Article number: 118128
ENVIRONMENT
- Ultrahydrogel
- Gel permeation chromatography (GPC)
- Size exclusion chromatography (SEC)
- Gel filtration chromatography (GFC)
ANSWER
No
ADDITIONAL INFORMATION
Deionized water is not recommended because it contains trace amounts of organic compounds that will contaminate the column.
Use MilliQ water instead, or other water that is not deionized water, such as HPLC-grade bottled water.
NOTE:
To avoid analyte adsorption to Ultrahydrogel packings, it is also advised to use an additive or buffer.
See the Waters Gel Permeation Chromatography (GPC) Capabilities Brochure for a table of recommended mobile phases for Ultrahydrogel columns.
Also see the Ultrahydrogel Columns Care & Use Manual.