What could cause my 30-mm preparative guard holder to leak from the weep hole? - WKB115305
Article number: 115305
ENVIRONMENT
- Column protection
- Preparative guard cartridges and holders
ANSWER
Several things could be causing this:
1. Cartridge not seated in the end fitting
To correct this problem, disassemble and properly seat the cartridge.
2. Particles on the face of the O-ring or metal seat
To correct this problem, wipe clean with a cotton-tip applicator.
3. Worn O-ring
Replace the worn O-ring with a new one (186007012).
ADDITIONAL INFORMATION
Care and use manual:
30mm ID Preparative Guard Cartridges and Holder Care & Use Manual (720004671en)
id115305,