Can I connect a nanoEase M/Z column directly into the Bruker nanoElute Valco port? - WKB115094
Article number: 115094
ENVIRONMENT
- nanoEase M/Z
- ZenFit
ANSWER
This kind of direct connection has not been tested by Waters.
ADDITIONAL INFORMATION
Because of dimensional tolerance differences, installing a nanoEase M/Z column (with ZenFit fittings) directly into a non-Waters V-detail (10-32, 1/16-inch OD tubing) port is not recommended.
Instead, connect a short piece of fitting and tubing between the port and the inlet of a ZenFit union (700011519). The inlet to the union connects with short tubing and fitting. Next, connect the nanoEase M/Z trap or column (with ZenFit fittings) to the outlet of the ZenFit union.