What are the suggested cleaning solutions for XBridge BEH Phenyl columns? - WKB109648

Article number: 109648

ENVIRONMENT

  • clean
  • solution
  • wash
  • XBridge
  • BEH
  • Phenyl
  • column
  • care 

ANSWER

If more polar samples were injected, the cleaning solvents are water, methanol, THF, methanol, water, and mobile phase.

If more nonpolar samples were injected, the cleaning solvents are IPA (or IPA:water mix), THF, DCM, hexane, IPA, and mobile phase.

ADDITIONAL INFORMATION

For additional information, see XBridge Columns Care and Use Manual Section V, Column Cleaning, Regeneration, and Storage or XBridge XP 2.5 µm Columns Care and Use Manual Section IV, Column Cleaning, Regeneration and Storage.

id109648, eluent, isopropanol