What is the difference between XBridge C18, Symmetry C18, and SunFire C18? - WKB108764
Article number: 108764
ENVIRONMENT
- Ultra High Performance Liquid Chromatography (UHPLC)
- High Performance Liquid Chromatography (HPLC)
- Column
- Bridged Ethylene Hybrid (BEH)
- High Performance Liquid Chromatography (HPLC)
- Column
- Symmetry
- Silica-based packing
- SunFire
ANSWER
XBridge BEH C18 packing is made with fully porous, 130A, Ethylene Bridged Hybrid (BEH) particles with tri-functionally bonded C18 ligands and is fully endcapped. The pH range for this packing is 1-12.
Symmetry C18 packing is made with fully porous, 100A, high-purity silica particles with mono-functionally bonded C18 ligands and is fully endcapped. The pH range of this packing is 2-8.
SunFire C18 packing is made with fully porous, 100A, high-purity silica particles with di-functionally bonded C18 ligands and is fully endcapped. The pH range of this packing is 2-8.