What are some method development starting points for Atlantis PREMIER BEH C18 AX mixed-mode, reverse-phase anion exchange columns? - WKB108534
Article number: 108534
ENVIRONMENT
- High Performance Liquid Chromatography (HPLC)
- Ultra High Performance Liquid Chromatography (UHPLC)
- High Performance Liquid Chromatography (HPLC)
- Method development
- Atlantis PREMIER BEH C18 AX mixed-mode, reverse-phase, anion exchange columns
- Polar compound retention
- VanGuard
ANSWER
Methods Development - Good places to Start
- Buffer Concentration
- Ammonium formate is a good choice for pH 2.7 – 4.7 and ammonium acetate for pH 3.8 – 5.8 (both are MS compatible)
- Start with 10 – 20 mM buffer
- Lower Concentration
- Retention increases for anions, decreases for cations
- Tailing for anions increases
- Higher Concentration
- Retention decreases for anions, increases for cations
- Improved peak shape for anions
- Lower Concentration
- Buffer pH
- Starting at low pH (3 - 4) tends to give sufficient retention for acids with the best peak shapes
- At higher pH (>6) the surface charge of the stationary phase gradually decreases
- Lower retention for anions, higher retention for cations
- Organic Content
- Increase the % organic for hydrophobic analytes
