What are some method development starting points for Atlantis PREMIER BEH C18 AX mixed-mode, reverse-phase anion exchange columns? - WKB108534

Article number: 108534

ENVIRONMENT

  • High Performance Liquid Chromatography (HPLC)
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • Method development
  • Atlantis PREMIER BEH C18 AX mixed-mode, reverse-phase, anion exchange columns
  • Polar compound retention
  • VanGuard

ANSWER

Methods Development  - Good places to Start

  • Buffer Concentration
    • Ammonium formate is a good choice for pH 2.7 – 4.7 and ammonium acetate for pH 3.8 – 5.8 (both are MS compatible)
    • Start with 10 – 20 mM buffer
      • Lower Concentration
        • Retention increases for anions, decreases for cations
        • Tailing for anions increases
      • Higher Concentration
        • Retention decreases for anions, increases for cations
        • Improved peak shape for anions
  • Buffer pH
    • Starting at low pH (3 - 4) tends to give sufficient retention for acids with the best peak shapes
    • At higher pH (>6) the surface charge of the stationary phase gradually decreases
      • Lower retention for anions, higher retention for cations
  • Organic Content
    • Increase the % organic for hydrophobic analytes