Are BEH C18 columns stable in a mobile phase of pH 10? - WKB107260

Article number: 107260

ENVIRONMENT

  • UltraPerformance Liquid Chromatography (UPLC)
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • Column
  • ACQUITY
  • Bridged Ethylene Hybrid (BEH)
  • C18
  • Column use
  • Column stability

ANSWER

Yes.

ADDITIONAL INFORMATION

BEH C18 packings are stable in a pH up to 12.0 with a maximum temperature of 60° C when in high pH mobile phases.

Also see: Waters Columns, Analytical Standards & Reagents Selection Guide Wall Chart

ACQUITY UPLC and ACQUITY PREMIER BEH Columns Care and Use Manual  

XBridge BEH XP 2.5um Columns Care & Use Manual

XBridge Columns Care & Use Manual