Can I use nanoEase m/z 150 um ID columns and nanoEase m/z 180 um ID trap columns on my nanoACQUITY system if I am using the nanoTee that has M-detail ports? - WKB100068

Article number: 100068

ENVIRONMENT

  • nanoEase m/z
  • nano
  • Column
  • Trap
  • nanoACQUITY
  • nanoTee

ANSWER

No.

ADDITIONAL INFORMATION

nanoEase m/z columns with 75 um, 100 um and 150 um ID with ZenFit fittings and nanoEase m/z trap columns in 180 um ID with ZenFit fittings, cannot screw into the M-detail ports (6-40 thread screw on 1/32" OD tubing) that are on the standard nanoTee used on nanoACQUITY systems.

Instead, use ACQUITY M-Class 75 um, 100 um and 150 um ID columns and ACQUITY M-Class 180 um ID trap columns on a nanoACQUITY system.