What are the amino acid sequences of the Waters standards 186008927, 186009125, and 186009126? - WKB96794
ENVIRONMENT
-
The following three Waters standards are produced from the same mAb protein. The NIST mAb (humanized IgG1k) monoclonal antibody is a recombinant antibody expressed in murine suspension culture.
-
Humanized mAb Mass Check Standard [186009125]
-
The sample consists of the intact mAb made of two light chains (LC) and two heavy chains (HC) linked with 16 disulfide bridges. The average molar mass should be 145179.8052 g/mol without modifications.
-
-
mAb Subunit Standard [186008927]
-
The sample consists of the Humanized mAb Mass Check Standard [186009125], which is reduced and digested with Ides. The Ides enzyme cleaves the heavy chain between the glycine-glycine amino acid residues. This sample consist of three sequences: Fc/2, Fd', and LC.
-
-
mAb Tryptic Digest Standard [186009126]
-
This sample consist of the Humanized mAb Mass Check Standard [186009125], which is reduced and alkylated to cleave the disulfide bridges. Then it is digested with trypsin. This sample consists of the two sequences HC and LC.
-
-
ANSWER
mAb Subunit Fc/2 sequence (Average molar mass: 23790.6425)
GPSVFLFPPK PKDTLMISRT PEVTCVVVDV SHEDPEVKFN WYVDGVEVHN AKTKPREEQY NSTYRVVSVL TVLHQDWLNG
KEYKCKVSNK ALPAPIEKTI SKAKGQPREP QVYTLPPSRE EMTKNQVSLT CLVKGFYPSD IAVEWESNGQ PENNYKTTPP
VLDSDGSFFL YSKLTVDKSR WQQGNVFSCS VMHEALHNHY TQKSLSLSPGK
mAb Subunit Fd' sequence (Average molar mass: 25705.8931)
QVTLRESGPA LVKPTQTLTL TCTFSGFSLS TAGMSVGWIR QPPGKALEWL ADIWWDDKKH YNPSLKDRLT ISKDTSKNQV VLKVTNMDPA
DTATYYCARD MIFNFYFDVW GQGTTVTVSS ASTKGPSVFP LAPSSKSTSG GTAALGCLVK DYFPEPVTVS WNSGALTSGV HTFPAVLQSS
GLYSLSSVVT VPSSSLGTQT YICNVNHKPS NTKVDKRVEP KSCDKTHTCP PCPAPELLG
mAb LC (Average molar mass: 23127.5093)
DIQMTQSPST LSASVGDRVT ITCSASSRVG YMHWYQQKPG KAPKLLIYDT SKLASGVPSR FSGSGSGTEF TLTISSLQPD DFATYYCFQG
SGYPFTFGGG TKVEIKRTVA APSVFIFPPS DEQLKSGTAS VVCLLNNFYP REAKVQWKVD NALQSGNSQE SVTEQDSKDS TYSLSSTLTL
SKADYEKHKV YACEVTHQGL SSPVTKSFNR GEC
mAb HC (Average molar mass: 49478.5203)
QVTLRESGPA LVKPTQTLTL TCTFSGFSLS TAGMSVGWIR QPPGKALEWL ADIWWDDKKH YNPSLKDRLT ISKDTSKNQV VLKVTNMDPA
DTATYYCARD MIFNFYFDVW GQGTTVTVSS ASTKGPSVFP LAPSSKSTSG GTAALGCLVK DYFPEPVTVS WNSGALTSGV HTFPAVLQSS
GLYSLSSVVT VPSSSLGTQT YICNVNHKPS NTKVDKRVEP KSCDKTHTCP PCPAPELLGG PSVFLFPPKP KDTLMISRTP EVTCVVVDVS
HEDPEVKFNW YVDGVEVHNA KTKPREEQYN STYRVVSVLT VLHQDWLNGK EYKCKVSNKA LPAPIEKTIS KAKGQPREPQ VYTLPPSREE
MTKNQVSLTC LVKGFYPSDI AVEWESNGQP ENNYKTTPPV LDSDGSFFLY SKLTVDKSRW QQGNVFSCSV MHEALHNHYT QKSLSLSPGK