Is RapiGest SF considered to be heat labile? - WKB73329
Article number: 73329
ENVIRONMENT
- RapiGest SF
- Solubilization factor (SF)
ANSWER
No
ADDITIONAL INFORMATION
RapiGest SF will not degrade in high temperatures (e.g., 95 °C), provided the pH of the solution is >6.0.
Lowering the pH to pH 2.0–3.0 initiates degradation of RapiGest.
Heating the low-pH solution further promotes the hydrolysis to form its products.
The half life of RapiGest SF in pH 2.0 is 8 minutes.
The half life of RapiGest SF in pH 3.0 is 60 minutes.
RapiGest SF Brochure (720000468en)
RapiGest SF Surfactant Care & Use Manual (715000122)