What is the gradient table that was used to generate data for application note 720006482EN (Application of Difluoroacetic Acid to Improve Optical and MS Performance in Peptide LC-UV/MS)?
ENVIRONMENT
- Application notes
- IonHance
- Difluoroaceitic acid (DFA)
- Gradient table
ANSWER
The Quaternary Solvent Manager (QSM) gradient used to generate the MassPREP Peptide data with 0.1% DFA:
Time (min) Flow Rate (mL/min) %A %B Curve
Initial 0.500 95 5 Initial
2.0 0.500 95 5 6
37.0 0.500 60 40 6
40.5 0.500 35 65 6
45.5 0.500 35 65 6
49.5 0.500 95 5 6
60.0 0.500 95 5 6