What could be a cause of inconsistent area counts when running AccQ•Tag Ultra? - WKB68830
Article number: 68830
ENVIRONMENT
- AccQ•Tag Ultra
- Amino Acid Analysis
ANSWER
- Double check your pH. For AccQ•Tag to work properly, you need to have your samples within the pH range of 8.2 to 10.
- Vortex, or mix the sample immediately after adding the derivatization reagent, any delay in time can cause an incomplete derivatization reaction, or lower area counts.