Late retention times in amino acid analysis on ACQUITY H-Class system using AccQ•Tag Ultra reagents - WKB268285
Article number: 268285
SYMPTOMS
- Retention times of all amino acids are later than expected
- AMQ peak retention time is approximately 3.0 minutes (should be approximately 1.5 minutes)
- Gradient pressure trace looks different from the normal gradient pressure trace when method was working
ENVIRONMENT
- Amino acid analysis (AAA)
- AccQ•Tag Ultra
- ACQUITY H-Class
CAUSE
Clogged/plugged GPV filters
System previously run with another method and buffer.
FIX or WORKAROUND
- Ensure that the correct mobile phases are in use for the method.
- Change out the GPV filters.
ADDITIONAL INFORMATION
id268285, 2695UP, eluent, gradient proportioning valve, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO