Can I increase the volume of my sample for AccQ•Tag derivatization of amino acids? - WKB219430
Article number: 219430
ENVIRONMENT
- Amino acid analysis (AAA)
- AccQ•Tag
- AccQ•Tag Ultra
- AccQ•Fluor
ANSWER
Yes
ADDITIONAL INFORMATION
The volume of the sample can be increased, but the borate buffer and AccQ•Tag reagent must also be increased to maintain the derivatization solution ratio of buffer and aqueous component of the mixture to organic content of the reconstituted AccQ•Tag reagent in the reaction vial.
For additional information, see Derivatization Guidelines in the Comprehensive Guide to Hydrolysis and Analysis of Amino Acids.