Skip to main content
Waters

High mass calibration fails with sodium iodide (NaI) solution - WKB202813

Article number: 202813

SYMPTOMS

  • Failure to create a calibration greater than m/z 2500 with sodium iodide (NaI) solution

ENVIRONMENT

  • IntelliStart
  • Assisted Calibration
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

The NaI solution is old, requires vortex mixing after storage, or the concentration is less than 2 mg/mL.

In negative ion mode, mass range includes m/z 400 or less.

FIX or WORKAROUND

  1. Try vortex mixing the old solution, If no improvement seen use a fresh 2-mg/mL NaI solution.
  2. Set the lower limit of negative ion mode to m/z 400.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.