How do I prepare the Cyclic IMS Trisaccharide Mix (186009273) for injection? - WKB200142

Article number: 200142

OBJECTIVE or GOAL

Reconstitute the Cyclic IMS Trisaccharide Mix (186009273) prior to injection.

ENVIRONMENT

  • Cyclic IMS Trisaccharide Mix

PROCEDURE

  1. Prepare a diluent made with 70:30 acetonitrile/water.
  2. Decant 10 mL of the diluent into a beaker.
  3. Dissolve the contents of the amber vial (1 mg of each sugar) in the 10 mL of diluent to make a stock standard.
  4. Dilute 1000x with 70:30 acetonitrile/water to make a solution for injection. 

ADDITIONAL INFORMATION

 