2695 Separations Module "lost prime" error - WKB82630

Article number: 82630

SYMPTOMS

  • Loses prime while running

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module

CAUSE

Insufficient back pressure

FIX or WORKAROUND

Add column or back pressure regulator to give more than 50 psi.

 

ADDITIONAL INFORMATION