2695 Separations Module "lost prime" error - WKB82630
Article number: 82630
SYMPTOMS
- Loses prime while running
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- 2795 Separations Module
CAUSE
Insufficient back pressure
FIX or WORKAROUND
Add column or back pressure regulator to give more than 50 psi.
ADDITIONAL INFORMATION